2-40
IM 701450-51E
2.8
Error Messages
A message may appear during operation. This section describes the meanings of the
messages and their corrective actions. This section lists only the error messages related
to the CAN Bus analysis function. There are other error messages related to the
DL7440/DL7480 and communications. These messages are described in the
DL7440/
DL7480 User’s Manual (IM701450-01E)
and the
DL7440/DL7480 Communication
Interface User’s Manual (IM701450-17E)
.
You can set the messages to be displayed in English or Japanese. For the procedure of
setting the message language, see section 15.1, “Changing the Message Language and
Turning ON/OFF the Click Sound” in the
DL7440/DL7480 User’s Manual (IM701450-
01E)
.
If the corrective action requires servicing, contact your nearest YOKOGAWA dealer for
repairs.
Code
Message
Action
Page
27
Executed the search, but no record was found
–
2-23, 2-32
that matched the pattern.
37
Aborted the analysis.
–
2-21
38
Data not detected.
–
2-5, 2-11,
Execute again after changing the settings or
2-19, 2-27
reacquiring the waveform.
39
The corresponding field was not found.
–
2-25
704
Cannot be executed while running.
Stop the acquisition.
Section 7.1 in
IM701450-01E
730
Pattern is not specified.
Select at least one field or frame type.
2-23, 2-32
Set at least one of the search pattern to a
value other than X.
739
Analyzed data does not exist.
Execute the analysis.
2-21, 2-30
Execute the analysis.
851
Computation can be carried out at the
Shorten the record length or turn OFF the
Section 7.2 in
current record length.
stuff bit computation.
IM701450-01E
2-20, 2-29
870
Cannot be specified.
Increase the number of valid bytes.
2-8, 2-15
Invalid byte or bit.