1-16
IM 701450-51E
• Activate a Trigger Only on the Address Pattern
Trigger Condition
Address Pattern:
A4
Data1 Pattern:
Not applicable
Condition:
True
Data2 Pattern:
Not applicable
Condition:
True
Ignore the Restart Condition:
No
Ignore Unexpected Start/Stop Condition:
No
<Byte Count=0>
S
25
AE
57
27
FE
98
99
27
P
A4
A R/W bit
Match against the specified address pattern.
Trigger activated here.
<Byte Count=3>
S
25
AE
57
27
FE
98
99
27
P
A4
3. Trigger activated here
1. Compare with the specified address pattern. Match
2. Skip 3 bytes
A R/W bit
• Activate a Trigger Only on the Data1 Pattern
Trigger Condition
Address Pattern:
Not applicable
Data1 Pattern:
27
Condition:
True
Data2 Pattern:
Not applicable
Condition:
True
Ignore the Restart Condition:
No
Ignore Unexpected Start/Stop Condition:
No
<Byte Count=0>
S
27
AE
57
27
FE
98
99
27
P
A4
2. Trigger activated here
1. Compare with the specified pattern (27). Match
A R/W bit
<Byte Count=3>
S
27
AE
57
99
FE
98
99
27
P
A4
3. Trigger activated here
1. Compare with the specified pattern (27). Match
2. Skip 3 bytes
A R/W bit
1.4 Setting the I
2
C Bus Signal Acquisition Conditions