C Bus Signal Analysis Function
1
Setting the Combination Trigger
22.
Press the Combination soft key. The Combination menu appears.
23.
Press the I
2
C Only, I
2
C on Pattern, or I
2
C -> Pattern soft key.
Select I
2
C Only to activate a trigger only on the trigger conditions of the I
2
C Bus signal.
Select I
2
C on Pattern to activate a trigger when the trigger conditions of the I
2
C Bus
signal are met while the trigger conditions of CH3 to CH8 (CH3 and CH4 on the
DL7440) are met. Select I
2
C -> Pattern to activate a trigger when the trigger conditions
of CH3 to CH8 (CH3 and CH4 on the DL7440) are met after the trigger conditions of
the I
2
C Bus signal are met.
When I
2
C Only Is Selected
24.
Press the Set Pattern soft key. The Set Pattern dialog box displays the
assignment conditions of the I
2
C Bus signal of CH1 and CH2.
25.
Press ESC. The Set Pattern dialog box closes.
Proceed to step 29.
1.4 Setting the I
2
C Bus Signal Acquisition Conditions