3-8
IM 701450-51E
When SPI Only Is Selected
20.
Press the Set Pattern soft key. The Set Pattern dialog box displays the
assignment conditions of the SPI Bus signals of CH1 to CH4.
21.
Use jog shuttle & SELECT to set the slope of the CH1 (SCK) edge to (rising
edge) or (falling edge).
22.
Use jog shuttle & SELECT to set the CH4 (SS) enable state to H (high) or L
(low).
Set the CH1 (SCK) edge slope
to or .
Set the CH4 (SS) enable state
to H or L.
23.
Press ESC. The Set Pattern dialog box closes.
Proceed to step 27.
When SPI on Pattern Is Selected
20.
Press the Set Pattern soft key. The Set Pattern dialog box displays the
assignment conditions of the SPI Bus signals of CH1 to CH4 and the pattern
setup screen of CH5 to CH8.
21.
Use jog shuttle & SELECT to set the slope of the CH1 (SCK) edge to (rising
edge) or (falling edge).
22.
Use jog shuttle & SELECT to set the CH4 (SS) enable state to H (high) or L
(low).
23.
Use jog shuttle & SELECT to set the patterns of CH5 to CH8 using H (high), L
(low), and X.
24.
Use jog shuttle & SELECT to set the pattern condition to True or False
(Condition box).
Set the CH1 (SCK) edge slope
to or .
Set the CH4 (SS) enable state
to H or L.
Select the CH5 to CH8 pattern
using H, L, and X.
Set the pattern condition to
True or False.
25.
Press ESC. The Set Pattern dialog box closes.
Proceed to step 27.
3.4 Setting SPI Bus Signal Acquisition Conditions