3-38
IM 701450-51E
Executing the Save Operation
8.
Press the Save Exec soft key. The data is saved to the directory indicated by
Path=...... At the same time, the Save Exec soft key changes to the Abort soft
key.
Aborting the Save Operation
9.
Press the Abort soft key. The save operation is aborted. At the same time, the
Abort soft key changes to the Save Exec soft key.
Specifying the Files to Be Displayed in the File List Window and Displaying
Properties
10.
Carry out steps 22 to 25 on page 12-23 in the
DL7440/DL7480 User’s Manual
(IM701450-01E).
Explanation
The analysis results of SPI signals can be saved to a file in ASCII format. The contents
of the detailed analysis list of the SPI signal are saved as-is to the file. The extension is
.txt.
The file size is as follows:
File size = (number of bytes per data point
1
×
number of analysis results) + 44 bytes
2
1 The number of bytes per data varies depending on the data type.
•
40 bytes (Data2) minimum.: Analysis data without CS
•
44 bytes (CS) maximum.:
Analysis data with multiple CSs and CS is set to CH4
2 The data size of the title is 44 bytes.
[Save example]
Precautions to Be Taken When Saving the Data
• The maximum number of files that can be saved when the auto naming function is ON
is 1150.
• If the total number of files and directories exceed 2500 in a single directory, the
contents of the File List box are no longer displayed.
3.6 Saving the Data of the Detailed Analysis List