1-37
IM 701450-51E
I
2
C Bus Signal Analysis Function
1
1.7
Error Messages
A message may appear during operation. This section describes the meanings of the
messages and their corrective actions. This section lists only the error messages related
to the I
2
C Bus signal analysis function. There are other error messages related to the
DL7440/DL7480 and communications. These messages are described in the
DL7440/
DL7480 User’s Manual (IM701450-01E) and the DL7440/DL7480 Communication
Interface User’s Manual (IM701450-17E).
You can set the messages to be displayed in English or Japanese. For the procedure of
setting the message language, see section 15.1, “Changing the Message Language and
Turning ON/OFF the Click Sound” in the
DL7440/DL7480 User’s Manual (IM701450-
01E).
If the corrective action requires servicing, contact your nearest YOKOGAWA dealer for
repairs.
Code
Message
Action
Page
37
Aborted the analysis.
–
1-27
38
Data not detected.
–
1-4, 1-10,
Execute again after changing the settings or
1-25, 1-30
reacquiring the waveform.
730
Pattern is not specified.
Enable Address Pattern or Data Pattern in the
1-29
Search Setup dialog box.
739
Analyzed data does not exist.
Execute the analysis.
1-27, 1-31
Execute the analysis.
870
Cannot be specified.
Increase the number of valid bytes.
1-6, 1-11
Invalid byte or bit.