3-14
IM AQ6373-01EN
Note
• Always use the reference light source built in to the instrument for alignment adjustments.
Alignment adjustments will be incorrect if an external light source is used.
• The alignment adjustment has no effect if the process is stopped. The instrument remains in
the state prior to execution of the alignment adjustment.
• Correct alignment adjustment may not be obtained with optical fibers other than 50/125 µm
GI.
3.5 Alignment Adjustment