No. Message
Cause of Warning
211 Auto offset error!
Obtained an abnormal value during AUTO OFFSET operation.
212 Auto offset error!
Obtained an abnormal value during AUTO OFFSET operation.
213 Auto temperature control error!
Problem with the light detector temperature control.
214 Measurement sequence error!
The sweep stopped because measurement sequence fell into
disorder during the sweep.
220 Boot sequence error!
Started up in emulation mode because an abnormality occurred
upon start-up
221 Boot sequence error!
Started up in emulation mode because an abnormality occurred
upon start-up
222 Emulation Mode.
Entered emulation mode because an abnormality occurred.
223 Boot sequence error!
Started up in emulation mode because an abnormality occurred
upon start-up
224 Internal communication error!
An abnormality occurred during instrument-internal communication.
225 Internal communication error!
An abnormality occurred during instrument-internal communication.
226 Internal communication error!
An abnormality occurred during instrument-internal communication.
227 Internal communication error!
An abnormality occurred during instrument-internal communication.
230 Monochromator error!
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
231 Monochromator error!
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
232 Monochromator error!
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
233 Monochromator error!
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
231 Monochromator error!
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
No.300 - 399: Errors during the execution of program functions
300 Parameter out of range
A variable value is out of range or is not defined for a command that
sets a parameter using variables.
302 Scale unit mismatch
There is a difference between the Y-axis scale of the active trace
and the unit of a parameter in the “LINE MKR 3 or 4” command.
303 No data in Active trace
Setting of the moving marker, a peak (or bottom) search, or
activation of the analysis function was made with no data in the
active trace.
304 Marker value out of range
Specified wavelength was out of the sweep range in the moving
marker or line wavelength marker setting command.
305 No data in trace A or B
No waveform data in traces A or B when executing the “EDFA NF”
command
306 Invalid data
Trace had no data when attempting to save it to memory or to write
it to FD/HDD.
307 Unsuitable Write item
All data items were OFF at execution of “WRITE DATA”.
320 Undefined variable
A command containing an undefined variable was executed.
321 Variable unit mismatch
The unit of each variable does not agree within a command
containing two or more variables.
322 Overflow
An overflow occurred in an arithmetic operation.
323 Undefined marker variable
A command containing a marker-value variable was executed when
no marker had been displayed.
324 Invalid marker variable
A command containing the corresponding variable was executed at
a time other than immediately after execution of a spectrum width
search, peak search, etc.
325 Undefined line number
GOTO command’s jumping destination is a number other than 1 to
200.
326 F1 greater than F2
F1>F2 when the “IF F1 ≤ @@@@@ ≤ F2” command was executed.
340 Printer paper empty
No printer paper
341 Printer head up
No print is made because the printer’s head-up lever is raised.
345 Option does not respond
No response from an external device.