
3-9
IM AQ6373-01EN
Preparing for Measurement
1
2
3
4
5
6
7
8
9
10
11
App
Index
Explanation
Screen when the instrument was not shut down
If the shutdown procedure was not performed after the previous session, the following
message appears after start up.
Failure to properly shut down the instrument can result in damage to the monochromator.
When turning OFF the power, always perform the shut down procedure.
Press any key to clear this message.
Turning the Power OFF
1.
Press the
POWER
switch on the front panel of the instrument. A shut down
confirmation message is displayed along with the
YES
and
NO
soft keys.
2.
Press the
YES
soft key. The message, “AQ6373 is shutting down. Please wait...”
appears, and shut-down begins. If you do not wish to shut down, press the
NO
soft key. The screen returns to the original soft key menu.
3.
After the
POWER
switch changes from green to orange, turn OFF the
MAIN
POWER
switch on the rear panel of the instrument.
CAUTION
Do not cut the power to the instrument with the
MAIN POWER
switch on the rear
panel when an operation is in progress. The operating system configuration file will
not be backed up, possibly resulting in malfunctions upon start up the next time the
instrument is turned ON. Always use the above procedure to shut down.
You can also shut down the instrument using panel keys and soft keys.
1.
Press
SYSTEM
.
2.
Press the
MORE
soft key three times.
The SYSTEM
4/4 screen is displayed.
3.
Press the
SHUT DOWN
soft key.
4.
Press the
YES
soft key. Shut down begins.
5.
After the
POWER
switch changes from green to orange, turn OFF the
MAIN
POWER
switch on the rear panel of the instrument.
3.3 Turning the Power ON/OFF