9-19
IM AQ6373-01EN
Other Operations
1
2
3
4
5
6
7
8
9
10
11
App
Index
Explanation
Automatic Offset
When the AUTO OFFSET key is set to ON, the offset of the internal amplifier circuit is
adjusted at time intervals of approximately 10 minutes. (Default: ON)
If the AUTO OFFSET key is OFF, the auto-offset adjustment operation is not performed.
When you switch from OFF to ON, the offset adjustment is executed immediately.
When switched to ON, the offset adjustment is performed during repeat sweeping when
sweeping reaches 100%.
During a single sweep, the offset adjustment is performed when the sweep is completed.
During offset adjustment, “ZEROING” is displayed in the lower left corner of the screen.
ZEROING. . .
Note
• During the offset process, the receiving of remote commands, execution of processes, and
various processes for sending of talker commands are suspended.
• For the calibration procedure, see section 3.6, “Wavelength Calibration.”
9.5 Other Settings