
7-42
IM AQ6373-01EN
Analysis in the Zoom Area
Power Measurement in the Zoom Area
Optical power can be measured by integrating the measured waveform level
measurements in the zoom area.
1.
Zoom in on the measured waveform. Set the range you want to measure to the
display scale. For the procedure on zooming in on the waveform, see section 6.1,
“Zooming In/Out on a Waveform.”
2.
Press the
MARKER
switch. The soft key menu regarding the markers appears.
3.
Press the
MORE 1/3
,
MORE 2/3
soft keys in succession.
4.
Press the
SEARCH/ANA ZOOM AREA
soft key, and select ON.
5.
Press
ANALYSIS
. The soft key menu for analyzing measured waveforms
appears.
6.
Press the
ANALYSIS 1
soft key. The analysis function selection menu is
displayed.
7.
Press the
POWER
soft key. Analysis is performed between display scales, and
the results are displayed in the data area.
Data area
Note
• When the zoom area search function is enabled,
is displayed in inverse video.
• If the SEARCH/ANA ZOOM AREA key is OFF, analysis is performed over the entire range of
the measurement scale.
• For details on the POWER analysis function, see appendix 3, “Details of Analysis
Functions.”
7.12 Specifying an Analysis Range