9-18
IM AQ6373-01EN
Turning Automatic Offset ON/OFF
You can specify whether to have the instrument perform automatic offset adjustment on
its internal amplifier circuit.
1.
Press
SYSTEM
.
2.
Press the
MORE 1/4
soft key. The MORE 2/4 screen is displayed.
3.
Press the
AUTO OFFSET SETTING
soft key.
4.
Press the
AUTO OFFSET OFF ON
soft key to select ON. The settings turns ON
and OFF each time you press the key.
5.
Press the
INTERVAL
soft key.
6.
Enter the value on a ten-key to set the offset interval of execution.
Note
• An offset interval of 10 minutes is recommended.
• If the AUTO OFFSET key is OFF, the offset can fluctuate over time, and the level axis
performance can degrade. Always have it turned ON.
• When the AUTO OFFSET key is set to ON,
is displayed at the bottom of the screen.
9.5 Other Settings