3.5 Alignment Adjustment
WARNING
The instrument has a built-in reference light source for alignment adjustments, and
infrared light is always being output from the optical output connector. Never look
into the optical output connector. Infrared light entering the eyes can cause severe
injury and loss of vision.
Procedure
1.
Turn the power to the instrument ON. Turn ON the
MAIN POWER
switch and
press the
POWER
switch. For instructions on turning the power ON and OFF, see
section 3.3.
2.
Use a 50/125
µ
m GI optical fiber to connect the instrument’s optical input
connector with the optical output connector. Length: Up to 10 m.
GI optical fiber
(50/125
µ
m)
AQ6373
3.
Press
SYSTEM
. The soft key menu is displayed.
4.
Press the
OPTICAL ALIGNMENT
soft key.
5.
Press the
EXECUTE
soft key. Alignment adjustment is automatically executed.
Adjustment finishes a few minutes thereafter, and you are returned to the original
screen.
6.
Press the
CANCEL
soft key to cancel alignment adjustment partway through the
process.