
6-18
IM AQ6373-01EN
6.6 Normalized Display
Procedure
1.
Press
TRACE
. The soft key menu for trace settings appears.
2.
Press the
ACTIVE TRACE
soft key. Traces A through G appear in soft keys.
3.
Press the Trace G soft key. The Trace G setting menu is displayed.
4.
Press the
VIEW G DISP/BLANK
soft key and select
DISP
. Each time you press
the soft key
DISP
switches to
BLANK
, or viceversa.
5.
Press the
CALCULATE G
soft key.
6.
Press the
NORMALIZE
soft key.
7.
To normalize Trace A press the
G=NORM A
soft key, to normalize Trace B press
G=NORM B
, and to normalize Trace G, press
G=NORM C
.
Example of waveform before normalization
Example of waveform after normalization
Note
• Normalization is not possible if the waveform level is too low.
• See section 6.5 for information about NORMALIZE.