Code Error Message
Description
Corrective Action
1014 Soft Version Error
There is a discrepancy between the
version of the firmware installed in
the frame controller and the firmware
installed in the module.
Update the firmware on the frame controller and
the module. Contact your local YOKOGAWA dealer
for the latest version of the firmware.
1015 Module Type Error
The frame controller can not detect the
module.
Update the firmware on the frame controller.
Contact your local YOKOGAWA dealer for the
latest version of the firmware.
Note
Error codes 1020 to 1099 are remote command errors. These error messages are described in
the
Communications Interface User’s Manual (IM 735101-17EN)
.
DFB-LD Module, FP-LD Module, and DUAL FP-LD Module
Code Error Message
Description
Corrective Action
1100 LS Lock
Laser output is locked.
Attach an interlock connector to the rear of the frame controller.
If the same message continues to appear, set Lock to Off on the
frame controller’s SYSTEM screen.
1101 Temperature Error The temperature inside the module is
too high.
Turn the power off, and confirm that the frame controller’s
inlet and exhaust holes are not blocked. After waiting for some
time, restart the instrument. If the same message continues to
appear, servicing is required.
1102 Update Error
A failure occurred during a firmware
update.
Restart the instrument.
If the same message continues to appear, servicing is required.
1104 Light Output Limit
Operation1
The optical output of the light source
(device 1) has exceeded its upper
limit.
Restart the instrument.
If the same message continues to appear, servicing is required.
1105 Memory Check
Error
An error occurred during the start-up
memory check.
Restart the instrument.
If the same message continues to appear, servicing is required.
1106 Thermistor
Temperature
Error1
The thermistor detected a
temperature error. (For laser light
source 1.)
Restart the instrument.
If the same message continues to appear, servicing is required.
1107 Thermistor
Temperature
Error2
The thermistor detected a
temperature error. (For laser light
source 2.)
Restart the instrument.
If the same message continues to appear, servicing is required.
1108 PEL Over current
Error1
The Peltier current is abnormal. (For
laser light source 1.)
Restart the instrument.
If the same message continues to appear, servicing is required.
1109 PEL Over current
Error2
The Peltier current is abnormal. (For
laser light source 2.)
Restart the instrument.
If the same message continues to appear, servicing is required.
1110 Temperature Limit
Error
The internal temperature is abnormal. Turn the power off and confirm that the frame controller’s inlet
and exhaust holes are not blocked. After waiting for some
time, restart the instrument. If the same message continues to
appear, servicing is required.
1111 Opt Lock Error
Laser output is locked. Optical output
cannot be turned on.
Unlock the laser output.
For details see section 4.1, “Troubleshooting.” If the same
message continues to appear, servicing is required.
1112 Light Output Limit
Operation 2
The optical output of the light source
has exceeded its upper limit.
Restart the instrument.
If the same message continues to appear, servicing is required.
4.2 Error Messages
4-8
IM 735101-01EN