3.
Turn the instrument on while holding down the top soft key.
The update screen appears, and the instrument begins updating the firmware automatically.
Top soft key
<< Update Firmware >>
Updating...
Boot Rev. : 1.01-00
FPGA1 Rev. : 0.06-00
Progress S . . . . . . . . -> E
A completion message appears at the bottom of the screen when the update finishes
correctly. After that, the frame controller starts automatically.
<< Update Firmware >>
Updating...
Boot Rev. : 1.01-00
FPGA1 Rev. : 0.06-00
Progress S . . . . . . . . . . . . . -> E
Flash (Firmware) Update Complete
You must reboot this machine
Note
You can disconnect the USB memory device when the frame controller starts to run.
Displaying the Version Information
1.
Press
SYSTEM
to display the SYSTEM screen.
2.
Use the arrow keys to select Soft Version View, and press the
Select
soft key or
ENTER
.
The SOFTWARE VERSION screen appears. This screen shows the version information for
the frame controller and all of the installed modules.
4.4 Updating the Firmware
4-20