Well-Ventilated Location
Inlet holes are located on the right and bottom of the instrument. There are also exhaust
holes on the rear side. To prevent internal overheating, allow for enough space around
the instrument (see the figure below), and do not block the inlet and exhaust holes.
5 cm or more
5 cm or more
10 cm or more
When connecting cables, allow for enough space, above and beyond the space shown in
the figure above, to carry out the procedure.
Ambient Temperature and Humidity
Ambient temperature
5 to 40°C
Ambient humidity
20 to 80% RH (no condensation)
Flat, Even Surface
Install the instrument with the correct orientation on a stable, horizontal surface. If the
instrument is installed in a horizontal position, rubber stoppers can be attached to the
feet at the rear of the instrument to prevent the instrument from sliding. One set of rubber
stoppers (two stoppers) are included with the instrument.
Rubber stoppers
A9088ZM
Feet at the rear of the instrument
2.2 Installing the Instrument
2-4
IM 735101-01EN