![YOKOGAWA AQ2200 Series User Manual Download Page 89](http://html.mh-extra.com/html/yokogawa/aq2200-series/aq2200-series_user-manual_911202089.webp)
3.2 Optical Sensor Modules
Analog Out (ANALOG OUT)
Analog Signal Output Terminal
This terminal transmits a voltage level that corresponds to the level of the optical signal
that is applied to the sensor.
Use cables that are 3 m or shorter in length.
Other than AQ2200-202
ANALOG OUT
Item
Specifications
Connector type
Mini plug
Output level
AUTO mode
Approx. 0 to 2 V, proportional to the measurement power (W) for
each power range
LINEAR mode
Approx. 0 to 2 V, proportional to the range set for measurement
power (W)
LOG mode
Approx. 0 to 2 V, proportional to the range set for measurement
power (dBm)
Output impedance
Approx. 1 k
Ω
AQ2200-202
ANALOG OUT 1
ANALOG OUT 2
Item
Specifications
Connector type
Mini plug
Output level
AUTO mode
Approx. 0 to 2 V or 0 to 5 V, proportional to the measurement power
(W) for each power range
LINEAR mode
Approx. 0 to 2 V or 0 to 5 V, proportional to the range set for
measurement power (W)
LOG mode
Approx. 0 to 2 V or 0 to 5 V, proportional to the range set for
measurement power (dBm)
TRIGGER mode
Approx. 0 to 2 V or 0 to 5 V, pulse output
Output impedance
Approx. 100
Ω
3-3
IM 735101-01EN
External Input and Output
1
2
3
4
5
Index