General Specifications
Item
Specifications
Operating environment
Ambient temperature: 5 to 40°C
Ambient humidity: 20 to 80% RH (no condensation)
Altitude: 2000 m or less
Storage environment
Ambient temperature: –20 to 60°C
Ambient humidity: 20 to 80% RH (no condensation)
Altitude: 3000 m or less
Recommended calibration period
1 year (including installed modules)
Rated supply voltage
100 to 240 VAC
Rated supply frequency
50/60 Hz
Maximum power consumption
AQ2211: 170 VA (including installed modules)
AQ2212: 580 VA (including installed modules)
Withstand voltage (between
the power supply and case)
1.5 kVAC for 1 minute
Insulation resistance (between
the power supply and case)
500 VDC, 10 MΩ or greater
External dimensions
AQ2211: 212 (W) × 132.5 (H) × 400 (D) mm (excluding protrusions)
AQ2212: 425 (W) × 132.5 (H) × 500 (D) mm (excluding protrusions)
Weight
AQ2211: approx. 6 kg (not including installed modules)
AQ2212: approx. 11 kg (not including installed modules)
Safety standard
Compliant standard EN 61010-1
Overvoltage category (installation category) II
1
Pollution degree 2
2
EN 60825-1 (Laser safety)
Emissions
Compliant standard EN 61326-1 class A
EN 55011 class A, group 1
EMC Regulatory Arrangement in Australia and New Zealand EN 55011 class A, group 1
Korea Electromagnetic Conformity Standard (
한국 전자파적합성기준
)
EN 61000-3-2
EN 61000-3-3
This is a class A instrument designed for an industrial environment. Operation of this
equipment in a residential area can cause radio interference, in which case users will
be required to correct the interference.
Cable conditions
• Interlock connector (REMOTE INTERLOCK)
Use a BNC coaxial cable that is 3 m in length or less.
• External trigger input terminal (TRIGGER IN)
Use a BNC coaxial cable that is 3 m in length or less.
• External trigger output terminal (TRIGGER OUT)
Use a BNC coaxial cable that is 3 m in length or less.
• USB port (type A)
Use a shielded USB cable that is 3 m in length or less.
• USB port (type B)
Use a shielded USB cable that is 3 m in length or less.
• Ethernet port
Use a category 7 Ethernet cable that is 3 m in length or less.
Immunity
Compliant standard EN 61326-1 Table 2 (for industrial locations)
Cable conditions
Same as the emission cable conditions.
Environmental
standard
Compliant Standard
3
EN50581 Monitoring and control instruments including industrial monitoring and control
instruments.
Complied (735101, 735102, 735125, 735122, 735131, 735133, 735141, 735143,
735142, 735162, 735163, AQ2200112, AQ2200131, AQ2200132, AQ220202, AQ220232,
AQ220242, AQ2200312, AQ2200332, AQ2200342)
1 The overvoltage category (installation category) is a value used to define the transient overvoltage condition and includes
the rated impulse withstand voltage. Overvoltage category II applies to electrical equipment that is powered through a fixed
installation, such as a switchboard.
2 The pollution degree refers to the degree of adhesion of a solid, liquid, or gas which deteriorates withstand voltage or surface
resistivity. Pollution degree 2 applies to normal indoor atmospheres (with only non-conductive pollution).
5-2
IM 735101-01EN
5.1 AQ2211/AQ2212 Frame Controller