4. Operation Method
External Output Terminal (optional)
Parameters
Temperature Output
(ANALOG)
The current (mA) corresponding to the measured temperature is
output.
Output temperature range: 0 to 160
℃
Output current: 4 to 20
m
A
Resistive load:
600Ω or bellow
Resolution:
±
2
℃
(
±
0.2mA)
Connection: M4 screw terminal
Time-up Output
(TIME UP)
It is output when operation is completed, including an abortion.
a-contact (relay contact)
Contact capacity: 250V AC, 1A (resistance load)
Connection: M4 screw terminal
Alarm Output
(ALARM)
It is output when an abnormality is detected. Refer to "Safety
Device and Error Code" in Page53.
a-contact (relay contact)
Contact capacity: 250V AC, 1A (resistance load)
Connection: M4 screw terminal
RS485 output
(
RS485
)
RS485 communication function
Temperature/current output table
Temperature (
℃
)
Output current (mA)
0
4
20
6
40
8
60
10
80
12
100
14
120
16
140
18
160
20