55
4. Operation Method
Operation course (instrument dry)
Follow the procedures below for the setting of instrument dry:
1. Turn on the earth leakage breaker (switch ON).
Turn on the earth leakage breaker, touch screen
displays the left interface for 3 secs, and then switch
over to the next interface.
2. The touch screen displays the operation mode
and program executed last time.
Fluid sterilize program: sterilize temp.126
℃
, sterilize
time 30min
Current temp. in chamber 26
℃,
Normal pressure (green character and icon)
Normal bottle (green character and icon)
Normal cover locking (green character and icon)
2.1 The touch screen displays the operation mode
and program executed last time.
Press Operation mode area, the background color
becomes yellow, switch over the interface
3. Mode selection interface
Fluid sterilize prog background color is blue that
means it is the operation mode executed last time.
3.1 Mode selection interface
Press Instrument dry prog one time,
the background
color becomes yellow, switch over the interface.