10
2. Before Using
Preparation before operation
(1)
Determine the installation site
If there is a bump on the floor, the casters may receive excessive load and get damaged. In
this case, lift and move carefully by four or more persons.
Install the equipment referring to 1 to 5 of "Installation precautions" in Page
Error! Bookmark
not defined.
and 7.
(2)
After the unit is placed in the desired
position, lock the stopper button of the
casters
Only the two casters on the front side of the
unit are equipped with a stopper.
Lock
Release
(3)
Connecting the power
・
Connect the power referring to 7 to 10 of " Installation precautions " in Page 8 and 9.
(4)
Open the cover
Turn on the electric leakage
breaker at the right side of
equipment and turn on the
power to open the cover.
The safety lock is released
and the cover can be opened.
・
Slide the lock lever on the
cover to the right and grasp
the handle of cover to open it.
Turn the lock lever to the left to close the cover.
Turn the lock lever to the right to open the cover.
电源接通前不能开锁
(5)
Pour water into the cooling
water box
・
Pour 2000ml water into the
cooling water box, 1-2mm
higher than the red indicator
line.
・
After pouring water, please
plug
tightly
the
silicone
stopper.
Keep
locked
before
power on
Silicone
stopper
(filler)
Water level
indicator
Water
box
drain valve
Water
box
outlet
Cylinder
drain valve