69
4. Operation Method
Shortcut function (preheating)
Follow the procedures below for the setting of preheating:
1. Turn on the earth leakage breaker (switch ON).
Turn on the earth leakage breaker, touch screen
displays the left interface for 3 secs, and then switch
over to the next interface.
2. The touch screen displays the operation mode
and program executed last time.
Sterilize & dry program: sterilize temp.126
℃
, sterilize
time 30min;
Dry temp. 150
℃
, dry time 30min
Current temp. in chamber 32
℃
Normal pressure (green character and icon)
Normal bottle (green character and icon)
Normal cover locking (green character and icon)
2.1 The touch screen displays the operation mode
and program executed last time.
According
to
the
aforementioned
operation
procedures of instruction manual, set the required
operation mode program, ie. corresponding temp. and
time; pressure, water tank and cover are normal
(green display). By default operate the program!
Press Preheating area, the background color
becomes yellow, switch over the interface
3. Preheating operation interface
Preheat OFF background color is blue that means the
operation mode executed last time.
3.1 Preheating operation interface
Press ON, the background color becomes yellow,
switch over the interface