12
2. Before Using
Preparation before operation
(10)
Set the sterile samples
Set the samples to the chamber, putting them into the attached sample basket.
Put the sample or sterilization bag into the chamber so they should not block or cover the
sensor inside the chamber, exhaust outlet and end connection to pressure gauge. If they
are blocked or covered, the vapor cannot be discharged and the equipment cannot be
operated correctly. Do not spill the samples when taking them out from/putting them into
the chamber. The failure in piping system, bad smell or dirt may result in.
In case liquid such as medicinal solution or medium is sterilized, the amount of liquid
should be 60% or less of the capacity of container. They may be boiled over if too much
quantity is supplied.
Widely open the opening of sterilization bag when used. If it is closed, the samples are
insufficiently sterilized.
(11)
Close the door before operation
Make sure to close the door of equipment before operation. If the door is not fully closed,
the cooling water box may be touched and may cause a burn injury. Do not open the door
during operation.
(12)
Close the cover
Make sure that no foreign objects exist
on the packing of the cover and its
contact area before closing the cover.
If any foreign object exists, the vapor
may leak from the inside.
Turn the lock lever to the left to close the cover.
Turn the lock lever to the right to open the cover.
电源接通前不能开锁
Fully close the cover and slide the lock lever on the cover to the left side. If it is closed
inappropriately, the vapor blows out from the inside, which may cause a burn injury.
Do not press the hook and operate the lock lever for purposes other than maintenance of
equipment.
(13)
Attach the droplet tray
Attach the droplet tray to the equipment
to prevent the water drops that are
made from vapor that generate during
air purge from dropping down from the
packing onto the samples or onto the
top board. Pour off the droplets inside
the tray periodically.
Droplet tray
Keep
locked
before
power on