2. Before Using
Preparation before operation
(6)
Set the attached heater baffle onto the
bottom surface inside the chamber
・
The heater baffle stabilizes the sterile
samples inside the chamber as well as
protects the heater and sensor. Make sure
to set it.
(7)
Close the drain valve
Close the drain valve at the bottom on the
left side face of main unit. Water leak
omlurs if not fully closed, which may
cause the burn injury or no-load (water)
operation.
Connect an appropriate hose (inner
diameter: 12mm) to the rear water outlet
and lead it to the draining site.
(8)
Pour water into the chamber
Before setting the sterile samples, pour
water into the chamber to the water level
gauge (notch) position.
Insufficient water may cause the no-load
(water) operation. Check the water level
every time before operation. Refill it
before the level becomes too low. Water
is required to be poured at dissolution
operation, as well as sterilizing operation.
When the water level lowers, the
equipm
ent detects an abnormality (“Er20”)
and cuts off the heater. Depending on
the conditions of equipment, however, the
detection requires too much time, which
may cause the heater deterioration.
Refill water before the water level
becomes too low.
Water QTY for models
6500
~
6600 ml
(9)
Replace the water in cylinder
・
Fill distilled or purified water to conduct sterilization! Do not use tap water or well water
directly!
・
Confirm the water quality in cylinder before putting samples. If discolored or turbid,
please clean the cylinder, and replace with new distilled or purified water!
Set the heater baffle
horizontally.
Close
Open
く
Water
outllet
Min.
Max.
Water level
gauge