Cleaning the ion block assembly
4-49
20. Remove the extraction cone handle insulator from the extraction cone
handle.
21. Remove the extraction cone seal from the ion block.
Cleaning the ion block components
Required materials
•
Chemical-resistant, powder-free gloves.
•
Appropriately sized glass vessels in which to completely immerse
components when cleaning. Use only glassware not previously cleaned
with surfactants.
•
HPLC-grade (or better) methanol.
•
HPLC-grade (or better) water.
•
Formic acid.
•
Ultrasonic bath.
•
Source of oil-free, inert gas (for example, nitrogen) for drying (air-drying
optional).
•
Wash bottle containing HPLC-grade (or better) 1:1 methanol/water.
•
Large beaker.
Extraction cone seal