C-8
External Connections
Connecting the Edwards oil-free roughing pump
The oil-free roughing pump is an alternative to the oil-filled roughing pump.
To connect the oil-filled roughing pump, see
“Connecting the Varian oil-filled
Required materials
•
Chemical-resistant, powder-free gloves
•
7-mm nut driver
•
Sharp knife
•
Elbow (included in the Waters Rotary Pump Kit)
•
NW25 tee (included in the installation kit)
•
NW25 center rings (included in the installation kit)
•
NW25 clamps (included in the installation kit)
•
NW40 center rings (included in the installation kit)
•
NW40 clamps (included in the installation kit)
•
NW25/NW40 adaptor (included in the installation kit)
TP03165
Inlet flange
Exhaust port flange