4-80
Maintenance Procedures
3. Remove the filter cover from the instrument.
4. Lift the filter, vertically, from the its slot in the instrument.
Tip:
If necessary, use the needle-nose files to grasp the filter.
5. Dispose of the filter.
6. Fit the new filter into the instrument.
7. Fit the filter cover to the instrument.
8. Tighten the thumbscrew on the filter cover.
9. Close the access door to the fluidics pump.
Filter cover
Filter