4-36
Maintenance Procedures
3. Use the combined 2.5-mm Allen wrench and cone extraction tool to
unscrew the 4 captive ion block assembly securing screws.
4. Remove the ion block assembly from the PEEK ion block support.
TP03130
Ion block assembly
securing screws
TP03130
Ion block assembly
PEEK ion block
support