B-2
Materials of Construction and Compliant Solvents
Preventing contamination
For information on preventing contamination, refer to
Controlling
Contamination in LC/MS Systems
(part number 715001307). Visit
www.waters.com.
Items exposed to solvent
The items that appear in the following table can be exposed to solvent. You
must evaluate the safety issues if the solvents used in your application differ
from the solvents typically used with these items. See
prepare mobile phases” on page B-3
for details about the most common
ingredients used to prepare mobile phases.
Items exposed to solvent
Item
Material
Autotune reservoirs
High-density polyethylene
Gas exhaust port
Stainless steel
Ion block assembly
Stainless steel and PEEK
Ion block screws
Gold-plated stainless steel
Ion block support
PEEK
Isolation valve
Stainless steel and PEEK
O-rings
Viton
®
or PTFE-encapsulated Viton
Probe adjuster bellows
Viton
Probe adjuster assembly
Anodized aluminium, glass filled
acetal, Xylan
®
, and stainless steel
Probe shaft
PEEK
Pumping block
Alochromed aluminium and Xylan
Solvent waste/leak management
Tygon tubing
Source enclosure
Alochromed aluminium
Source enclosure view port
Toughened plate glass
Trap bottle
Polypropylene