4-18
Maintenance Procedures
Adding oil to the Varian roughing pump
If the roughing pump oil level is found to be low, you must add oil to the
roughing pump. See
“Inspecting the Varian roughing pump oil level” on
.
Varian roughing pump
Required materials
•
Chemical-resistant, powder-free gloves
•
Funnel
•
Varian GP45 oil
TP03159
Oil filler plug
Oil level sight glass
Oil drain plug