4-34
Maintenance Procedures
2. Hold the sampling cone assembly so that the cone gas nozzle handle is
oriented horizontally and at the top, and then slide the sampling cone
assembly into the ion block assembly.
3. Grasp the cone gas nozzle handle and use it to rotate the sampling cone
assembly 90 degrees, moving the handle downward from the horizontal
to the vertical position.
4. Open the source isolation valve (see
“To open the source isolation valve
after completing a maintenance procedure” on page 4-13
).
5. Close the source enclosure.
TP03132
Ion block assembly
Sampling cone assembly