
Replacing the source assembly seals
4-75
Replacing the source assembly seals
To avoid excessive leakage of solvent vapor into the laboratory atmosphere,
the following seals must be renewed at intervals of no greater than 1 year:
•
Probe adjuster assembly probe seal
•
Probe adjuster assembly nebulization gas seal
•
Source enclosure seal
•
Source enclosure nebulizer gas seal
•
Source enclosure desolvation gas seal
•
Source enclosure pressure test gas seal
Removing the probe adjuster assembly probe and source
enclosure seals
Required materials
•
Chemical-resistant, powder-free gloves
•
O-ring removal kit
Warning:
To avoid excessive leakage of solvent vapor into the
laboratory atmosphere, the seals listed below must be renewed,
at intervals of no greater than 1 year, exactly as described in this
section.
Summary of Contents for Xevo TQ MS
Page 10: ...x...
Page 42: ...2 10 Preparing the Mass Spectrometer for Operation...
Page 129: ...Replacing the air filter 4 81...
Page 130: ...4 82 Maintenance Procedures...
Page 146: ...B 4 Materials of Construction and Compliant Solvents...
Page 178: ...D 4 Plumbing the IntelliStart Fluidics system...