Cleaning the source components
4-25
Cleaning the source components
Clean the sample cone and cone gas nozzle when these conditions apply:
•
The sample cone and cone gas nozzle are visibly fouled.
•
LC and sample-related causes for decreased signal intensity have been
dismissed.
“Cleaning the sampling cone assembly” on page 4-26
.
If cleaning the sample cone and cone gas nozzles fails to increase signal
sensitivity, also clean the extraction cone. See
“Cleaning the extraction cone”
.
If cleaning the extraction cone fails to increase signal sensitivity, clean the ion
block and isolation valve. See
“Cleaning the ion block assembly” on page 4-42
.
If cleaning the ion block and isolation valve fails to increase signal sensitivity,
also clean the source hexapole assembly. See