Cleaning the instrument case
4-15
Cleaning the instrument case
Use a soft cloth, dampened with water, to clean the outside surfaces of the
mass spectrometer.
Emptying the nitrogen exhaust trap bottle
Inspect the nitrogen exhaust trap bottle in the instrument’s exhaust line,
daily and empty it before it is more than approximately 10% full.
Nitrogen exhaust trap bottle
Caution:
Do not use abrasives or solvents to clean the instrument’s
case.
TP03164
From instrument
exhaust connection
To laboratory
exhaust port
Cap
Nitrogen exhaust
trap bottle
From instrument
pilot valve port