4-66
Maintenance Procedures
6. Remove the metal gasket from the probe tip.
7. Remove the PEEK union/UNF coupling assembly and capillary from the
probe.
8. Unscrew and remove the knurled collar from the UNF coupling.
9. Remove the knurled collar and conductive sleeve from the capillary.
Metal gasket
PEEK union/UNF coupling assembly
Capillary
UNF coupling
PEEK union
Conductive sleeve
Knurled collar
Locknut