Maintenance
94
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
3
Open the chamber door.
4
Remove the dust filter and move the sterilizer closer to the shelf edge.
5
Push the reset button of the thermostat switch: a click sound indicates that the
thermostat switch has been reset.