LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
119
Technical data
Usable space *
LARA 17: 11.5 l/W: 195 mm/H: 195 mm/D: 297 mm (3 gal/W:
7.7’’/H: 7.7’’/D: 11.7’’)
LARA 22: 15 l/W: 195 mm/H: 195 mm/D: 390 mm (4 gal/W:
7.7’’/H: 7.7’’/D: 15.4’’)
Bacteriological filter
0.3 µm
Note*: Usable space with standard rack and trays. With optional tracks and
trays, see "Accessories, spare parts, consumables" on page 123.
STEAM GENERATOR
Pressure safety valve
5 bar (72.51 psig)
Safety thermostats
230 °C (446 °F)
DISTILLED OR DEMINERALIZED WATER
Water quality
Fulfilling EN 13060 Ann. C (conductivity < 15μS/cm)
Average water consumption
0.13 to 0.59 litres/cycle (0.03 to 0.16 gal/cycle)
Tank volume
Clean water 4.9 l (1.29 gal), with air gap: 2.8 l, (0.74 gal)
Used water 4.8 l (1.27 gal)