Sterilizer tests
46
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
PURPOSE OF THE TEST
The test is used to validate the sterilizer performance for textile load sterilization.
DESCRIPTION
It consists of several sheets of paper wrapped in a small packet with a chemical heat-sensitive indicator sheet in the middle. The colour
assumed by this indicator sheet at the end of the sterilization cycle gives the result of the test.
CARRY OUT THE TEST (IN ACCORDANCE WITH EN13060)
1
Empty the sterilizer chamber to ensure no load is present. Do not remove the
standard chamber accessories mounted (e.g. chamber rack and trays).
2
Place the Bowie and Dick test pack in the center of a tray in the lowest rack
position and close the chamber door.