64
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
Sterilization cycles
2
Tap
to abort the stop command.
The cycle continues programmed.
3
Tap
to abort the cycle: the
sterilizer starts a reset phase.
Notice: Do not switch off the sterilizer
during the reset phase: it takes some
time to reset the system and reach
safe conditions in the sterilizer
chamber.
4
Check the message. See
"Messages of a stopped
sterilization cycle" on the next
page.
5
Tap
to view the cycle
parameters. See "View the cycle
parameters" on page 62.
6
Open the chamber door.
7
Reprocess the load if necessary.
CAUTION! Hot steam. Wait
the steam to dissipate
before opening the door.
Note: Water could be present in the
chamber when opening the door. To
prevent spilling place a towel below the
chamber door.