Maintenance
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
81
REPLACING THE DUST FILTER
Notice: If you replace this consumable before the maintenance due date you have to reset the cycle counter. See "Replace the consumable
before the maintenance due date" on page 74.
Proceed as follows:
1
Open the chamber door.
2
Pull out the dust filter handle from underneath the sterilizer.
3
Lift the used filter from the handle and remove it.