Maintenance
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
89
800-cycle maintenance
REPLACING THE DOOR SEAL
Notice: If you replace this consumable before the maintenance due date you have to reset the cycle counter. See "Replace the consumable
before the maintenance due date" on page 74.
Proceed as follows:
1
Fully open the chamber door.
2
Remove the used door seal by hand.
3
Carefully clean the seal seat and the chamber face side with a damp, lint-free
cloth.
4
Moisten the new seal with water to facilitate its placement.