Maintenance
90
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
5
Insert the new seal and press first up and down.
6
Press left and right and then the entire seal circumference to ensure its perfect
placement.
Notice: A steam discharge can damage the sterilizer plastic parts. Ensure the
seal does not stick out.
7
Wipe any residual water and run a Vacuum and Helix test to check for perfect
tightness of the seal. See "Vacuum test" on page 52 and "Helix test" on page 49.