Diagnostic
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
103
Code
Description and actions
Actions
3xx
Check the door seal. Clean or replace it if necessary.
Repeat the cycle.
If the problem persists, call service.
Clean the chamber face side.
Clean the chamber filter.
Check if chamber filter is properly locked in the cap.
Check the load does not exceed the MAXIMUM WEIGHT
LIMITS.
4xx
Clean water error (bad quality or low water level). Drain and
refill the clean water tank.
Repeat the cycle.
If the problem persists, call service.
5xx
Check if there are hurdles on the door locking area
(chamber rack, loads, objects, …).
Repeat the cycle.
If the problem persists, call service.
Check the door seal (wrong placed).
Check if the door can move freely without touching the
trays or the load when closing.
Switch the sterilizer OFF and ON.
990
The cycle has been aborted by the user.
Re-process the load.
MESSAGES AND ALERTS
Notice: For any error not listed in this table, call technical service.
Message/Alert
Description
Action
Fill clean water tank.
There is not enough water in the tank to perform a
cycle.
Fill the water tank as requested.
Drain used water tank.
The used water tank is full.
Drain the water tank as requested.