Sterilizer tests
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
53
3
On the homepage tap Vacuum test.
4
Tap
and enter your credentials if required: the chamber door locks.
5
Wait until the end of the test and tap OPEN: the chamber door unlocks.
6
Enter your credentials if required: a message informs if the test passed or failed.
If the test failed, see "What to do when the test failed" below
WHAT TO DO WHEN THE TEST FAILED
1
Check, clean or replace the door gasket.
2
Clean the chamber face side and the chamber filter.
3
Repeat the Vacuum test. See "Carry out the test" on the previous page.
4
If the test fails repeatedly, call technical service.