
Sterilizer tests
50
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
4
Place the tube with the capsule in the center of a tray in the lowest rack position
and close the chamber door.
5
On the homepage tap Helix-B&D test.
6
To set the duration of the Plateau/Sterilization phase and other settings, tap
.
7
Tap
and enter your credentials if required: the chamber door locks.
8
Wait until the end of the test and tap OPEN: the chamber door unlocks.
9
Enter your credentials if required.