Sterilizer tests
48
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
INTERPRET THE TEST RESULT
Indicator
What happened
Test
passed
What to do next
The entire surface of the indicator sheet has changed colour.
Yes
-
Certain ares of the indicator sheet have not changed colour since there was an air
pocket during the cycle due to sterilizer malfunction.
No
Repeat the test. If it fails
repeatedly, call technical service.