Diagnostic
104
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
Message/Alert
Description
Action
Please close the door.
The door must be locked, but you didn't close it.
Close the door so it can be locked.
Non-conform water
The clean water quality is bad (conductivity
between 15 and 50 μS/cm).
You may run a cycle but the water must be replaced soon, otherwise the unit will
automatically lock-out to prevent damage.
Unacceptable water
The clean water quality is very bad (conductivity
more than 50 μS/cm).
Running a cycle is inhibited to prevent damage. Replace the clean water.
Door Gasket must be replaced in
... cycles.
Have you ordered the Door
Gasket?
These are pre-alerts advising that one of the
consumables has to be replaced within a small
number of cycles.
Tap
if you have the consumable available for replacement.
Tap
if you don't have the consumable in stock and must order one. In this case, the
pre-alert will appear again after some cycles.
Bact. Filter must be replaced in ...
cycles.
Have you ordered the Bact. filter?
Dust Filter must be replaced in ...
cycles.
Have you ordered the Dust filter?
4000 cycle maintenance must be
performed in ... cycles.
Have you already booked the
4000 cycle service?
This pre-alert advises you that the target of 4000
cycles is close and the relevant maintenance step
is to be scheduled.
Contact your Technical Service.
Possible leak detected.
Please run Vacuum Test.
Air was detected in the chamber: a vacuum leak is
suspected. The cycle was completed but a vacuum
test is required.
Run a vacuum test.
Call for service if an anomaly is detected.
Remote data storage
Please check settings.
Error when saving the files in the network.
Check the settings, the network resource and the memory space available on it.