Diagnostic
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
101
Code
Description and actions
Actions
0xx
Load cannot be considered sterile. See "End of a
sterilization cycle" on page 63
Repeat the cycle.
If the problem persists, call service.
Check if the mains switch or network circuit breaker is OFF.
Check if the mains cable is properly connected.
Switch the sterilizer OFF and ON.
Set date and time, then switch the sterilizer OFF and ON.
Check the dust filter and ensure that the sterilizer fan is not
blocked.
10x
See error "13x to 16x" on the next page.
Repeat the cycle.
If the problem persists, call service.
12x
Wait before opening the chamber door. Allow the
sterilization chamber to cool down. See error "13x to 16x" on
the next page.
Repeat the cycle.
If the problem persists, call service.