Diagnostic
LARA | Instructions for use | RIL-1 ENG Rev04 | 28/02/2019 | © 2019 W&H Sterilization Srl
105
Message/Alert
Description
Action
Remote data storage
File lost.
Some cycle reports couldn't be saved in the
network because the sterilizer memory is limited
(see "Sterilization cycle report" on page 65).
No action possible. The cycles were overwritten in the sterilizer database before being
saved, and there is no way to recover them.
Door Gasket replacement is due.
Have you replaced the door
gasket?
These messages advise that one consumable
must be replaced.
Replace the consumable and tap
to reset the counter (See "Maintenance" on
page 72).
If you don't replace the consumable, press
.
In this case, you may still use the sterilizer but the message will appear again after
some cycles.
CAUTION! Operating the sterilizer with expired consumables could be
dangerous and could damage the sterilizer.
Bact. Filter replacement is due..
Have you replaced the
bacteriological filter?
Dust Filter replacement is due.
Have you replaced the dust filter?
4000 cycle service
Please call your technical support
to organize service.
This message advises you that the target of 4000
cycles has been achieved and the relevant
maintenance step is to be performed.
Call technical service for the 4000 cycle maintenance.